BBD203 Connection Guide

Quick Start - Connecting Your BBD203 Controller

This guide explains how to connect your ThorLabs BBD203 motor controller to nueScan using the simplified serial number method.


Step 1: Physical Connection

  1. Connect USB Cable

    • Connect the USB cable from your BBD203 controller to your computer
    • Wait for Windows/Linux to recognize the device
    • No special drivers needed (uses standard FTDI USB-Serial)
  2. Power On Controller

    • Ensure BBD203 is powered on
    • Front panel should be lit

Step 2: Find Serial Number

The serial number is printed on a label on your BBD203 controller.

Common Locations:

  • Back panel of the controller
  • Side panel
  • Original packaging

Format:

  • Usually 8 digits (e.g., 83123456)
  • May include letters (e.g., 83A12345)

Example Label:

ThorLabs BBD203
S/N: 83123456

Step 3: Connect in nueScan

Using the GUI

  1. Launch nueScan

    python -m nuescan
    
  2. Enter Serial Number

    • Locate the "ThorLABS MLS Stage Serial:" field at the top of the window
    • Type your serial number (e.g., 83123456)
  3. Click Connect

    • Click the "Connect" button next to the serial field
    • Wait 1-2 seconds for connection
  4. Success!

    • If successful, you'll see a confirmation dialog
    • Button changes to "Disconnect"
    • All 3 motor channels are now enabled

Programmatic Connection

from hardware.thorlabs_stage import ThorLabsStage

# Create stage instance
stage = ThorLabsStage(encoder_counts_per_mm=20000)

# Connect by serial number
success = stage.connect('83123456')

if success:
    print("Connected! Ready to home axes.")
else:
    print("Connection failed.")

Troubleshooting

"Device not found" Error

Problem: Connection fails with "Could not find BBD203 with serial number..."

Solutions:

  1. Check USB Connection

    • Ensure USB cable is fully inserted
    • Try a different USB port
    • Try a different USB cable
  2. Verify Serial Number

    • Double-check the serial number on the controller label
    • Ensure no typos (0 vs O, 1 vs I, etc.)
  3. List Available Devices

    • The error dialog will show all detected ThorLabs devices
    • Check if your device appears with a different serial number
    • If no devices shown, check USB connection and drivers
  4. Windows: Check Device Manager

    • Open Device Manager
    • Look under "Ports (COM & LPT)"
    • Should see "USB Serial Port (COMx)" with FTDI in description
    • If device shows with "!" icon, driver issue
  5. Linux: Check Permissions

    # Check if device is detected
    lsusb | grep -i ftdi
    
    # Check serial ports
    ls -l /dev/ttyUSB*
    
    # Add user to dialout group (may require logout)
    sudo usermod -a -G dialout $USER
    

Connection Succeeds but No Response

Problem: Connection successful but motors don't respond

Solutions:

  1. Check Power

    • Verify motors are connected and powered
    • Check motor power LEDs on BBD203 front panel
  2. Enable Channels

    • Channels should auto-enable on connection
    • Check status indicators in UI
  3. Home Axes

    • Axes may need homing before movement
    • Try homing each axis

Multiple Controllers

Problem: You have multiple BBD203 controllers connected

Solution:

  • Each controller has a unique serial number
  • Connect to specific controller by entering its serial number
  • Error dialog will show all connected devices

What Happens During Connection

Automatic Process

When you click "Connect", the following happens automatically:

  1. USB Enumeration

    • Scans all USB ports
    • Finds ThorLabs devices (FTDI vendor ID: 0x0403)
    • Matches your serial number
  2. Port Assignment

    • Determines the COM port (e.g., COM3, /dev/ttyUSB0)
    • Opens serial connection at 115200 baud
  3. Controller Initialization

    • Requests hardware information
    • Enables automatic status updates
    • Enables all 3 motor channels
  4. Default Configuration

    • Sets velocity: 1.0 mm/s
    • Sets acceleration: 5.0 mm/s²
    • Starts position monitoring

Status Updates

After connection:

  • Position updates received every ~100ms
  • Status bits monitored (homed, moving, errors)
  • Move completion notifications enabled

Advanced Configuration

Custom Encoder Resolution

If you're using a stage with different encoder resolution:

# Example: Stage with 2,000 counts/mm instead of default 20,000
stage = ThorLabsStage(encoder_counts_per_mm=2000)
stage.connect('83123456')

Common resolutions:

  • MLS203: 20,000 counts/mm (default)
  • DDS220: 2,000 counts/mm
  • Custom: Check your stage specifications

If you need to connect to a specific port instead of using serial number:

from hardware.bbd203_driver import BBD203Driver

driver = BBD203Driver()
driver.connect('COM3')  # or '/dev/ttyUSB0' on Linux

System Requirements

Operating Systems

  • ✅ Windows 7/8/10/11
  • ✅ Linux (Ubuntu, Fedora, etc.)
  • ✅ macOS (with FTDI driver)

Dependencies

  • Python 3.8+
  • PySerial 3.5+
  • PyQt6 6.4+

USB Requirements

  • USB 2.0 or higher
  • FTDI USB-Serial drivers (usually automatic)

Next Steps

After successful connection:

  1. Home the Axes

    • Required before first movement
    • Establishes zero position reference
  2. Test Movement

    • Try small movements to verify operation
    • Check position feedback in UI
  3. Configure Scan Parameters

    • Set scan area (X/Y start, delta)
    • Set row spacing
    • Configure velocity if needed
  4. Begin Scanning

    • All systems should show "Ready"
    • Click "Begin Scan" to start

Support

For issues with the BBD203 driver or connection:

  1. Check Debug Output

    • Console window shows connection details
    • Look for "INFO:", "DEBUG:", and "ERROR:" messages
  2. Review Driver Documentation

    • BBD203_DRIVER_README.md - Complete driver reference
    • BBD203_Communications_Protocol.md - Protocol details
  3. ThorLabs Support


Example Session

$ python -m nuescan

# UI appears
# Enter serial: 83123456
# Click Connect

# Console output:
INFO: Connecting to BBD203 with serial number 83123456
DEBUG: Found ThorLabs device - Serial: 83123456, Port: COM3
INFO: Found device 83123456 on port COM3
INFO: Connecting to BBD203 on COM3
INFO: Successfully connected to BBD203 on COM3
INFO: Stage connected and channels enabled

# Success dialog appears
# Button changes to "Disconnect"
# Ready to home and move!

For detailed technical information, see BBD203_DRIVER_README.md