# Session summary — scan timeout & geometry fixes ## Bug fixed: oscilloscope timeout on first row **File:** `hardware/tektronix_base.py` — `transfer_fastframe()` (~line 1414) The scope was configured for N frames but only triggered on fewer (e.g. 57 of 58) because the stage decelerates before the last laser pulse. `transfer_fastframe` looped using `get_fastframe_count()` (the configured maximum), so the final `read_raw()` call blocked waiting for data that never came and timed out. **Fix:** replaced `get_fastframe_count()` with a live query: ```python acquired = int(self.query("ACQuire:NUMFRAMESACQuired?")) ``` The loop now reads exactly as many frames as the scope actually captured. --- ## Scan geometry overhaul **File:** `sc3_aui_app.py` ### Constants added (near line 71) ```python LASER_FREQ_HZ = 2000.0 # fixed laser pulse frequency SCAN_RAMP_MM = v² / (2a) # ≈ 3.33 mm (100² / 2×1500) ``` `SCAN_RAMP_MM` is the distance the stage needs to accelerate from rest to full scan velocity, or decelerate back to rest. ### `_run_scan()` changes - `points_per_row` now uses `LASER_FREQ_HZ` instead of the param-supplied `laser_freq`. - Pre-scan X position is `x_start - SCAN_RAMP_MM` so the stage arrives at `x_start` already at full velocity (TRIGOUT_MAXV fires at the right place). - Scan move ends at `x_start + x_delta + SCAN_RAMP_MM` so the stage doesn't begin decelerating until after the last data point. ### Travel-limit guards added (before any hardware interaction) Raises `ValueError` with an actionable message if the extended move would exceed the stage limits baked into the driver (`bbd20x.py`: X 0–110 mm, Y 0–75 mm): - `x_start - SCAN_RAMP_MM < 0` - `x_start + x_delta + SCAN_RAMP_MM > 110` - any Y row position outside 0–75 mm Error messages tell the user exactly how many mm to adjust. --- ## What to check / next steps - Verify `ACQuire:NUMFRAMESACQuired?` is the correct query string for the specific scope model in use (MDO/MSO series assumed; confirm against programmer manual). - `LASER_FREQ_HZ = 2000.0` is a temporary constant — wire it to the UI param when ready. - Confirm `SCAN_RAMP_MM` matches observed stage behaviour; if the BBD202 velocity profile is not perfectly triangular the empirical ramp may differ slightly from v²/2a.