Helios: read CRLF replies as lines, not as CR plus dead air

The rig transcript (tools/helios_lds_probe.py) settles where the panel's
32 mA came from, and it was never the laser: LDS reads 100 mA, answers for
itself, and takes a write of 900 mA on the first attempt. 32 is LCE — bit
5, "Door switch open" — arriving in the diode-current field.

Every reply is CRLF-terminated and padded with a blank line or two:

    b'LDS =    100 mA\r\n\r\n'
    b'LCE =     32\r\nBit 15..0: 0000 0000 0010 0000\r\n\r\n\r\n'

_read_line() read up to CR, so the final LF of every reply stayed in the
buffer, and the next read waited out the whole port timeout for a CR that
only the next command would bring. A second of dead air per query: replayed
against the transcript's byte timing, one status poll took 8.6 s against
the 1 s interval that schedules it. That is also what let the values drift
apart — a query whose deadline goes to a blocked read gives up while its
own reply is still on the wire, the next query flushes the port mid-line,
and the fragment it reads is "     32", the value half of LCE's reply.

Lines are now framed on CR, LF or CRLF out of a receive buffer that
_discard_input() clears along with the port, so nothing survives a flush
half-read. The same replay now polls in 0.59 s.

Tests carry the transcript's real framing (padded values, trailing blank
lines) instead of the tidied "LER = 0" it was guessed to be, plus the two
regressions: a late fragment must not become the next query's value, and a
reply must be readable without waiting out the port.

Co-Authored-By: Claude Opus 5 <noreply@anthropic.com>
This commit is contained in:
Thomas Ales
2026-09-08 08:01:27 -05:00
parent eff589d901
commit 76ed7828cc
3 changed files with 138 additions and 26 deletions
+48 -8
View File
@@ -34,6 +34,7 @@ class HeliosLaser:
self.timeout = timeout self.timeout = timeout
self.serial = None self.serial = None
self.is_connected = False self.is_connected = False
self._rx = bytearray() # bytes read off the port, not yet a line
@staticmethod @staticmethod
def list_available_ports() -> List[str]: def list_available_ports() -> List[str]:
@@ -52,6 +53,7 @@ class HeliosLaser:
try: try:
self.serial = open_8n1(self.port, baudrate=9600, timeout=self.timeout) self.serial = open_8n1(self.port, baudrate=9600, timeout=self.timeout)
self._rx.clear()
time.sleep(0.1) # Allow time for connection to stabilize time.sleep(0.1) # Allow time for connection to stabilize
self.is_connected = True self.is_connected = True
logger.info(f"Connected to Helios laser on {self.port}") logger.info(f"Connected to Helios laser on {self.port}")
@@ -108,20 +110,58 @@ class HeliosLaser:
TRAILING_QUIET_S = 0.05 # the line counts as idle after this long TRAILING_QUIET_S = 0.05 # the line counts as idle after this long
MAX_REPLY_LINES = 8 MAX_REPLY_LINES = 8
def _discard_input(self):
"""Drop anything unread, on the wire and already taken off it."""
self._rx.clear()
self.serial.reset_input_buffer()
def _read_line(self) -> Optional[str]: def _read_line(self) -> Optional[str]:
"""One CR-terminated line without its framing; None if nothing came.""" """One line, however it is framed; None if nothing came in time.
raw = self.serial.read_until(b'\r')
if not raw: The controller ends every line with CRLF and pads a reply with a
return None blank line or two:
return raw.decode('ascii', errors='replace').strip()
b'LDS = 100 mA\r\n\r\n'
Reading up to CR alone leaves the trailing LF behind, and the next
read then waits out the whole port timeout for a CR that will not
come until some later command is answered. That was a second of
dead air per query — a status poll took ~8.6 s against the 1 s
interval that schedules it — and worse, a query that spends its
deadline blocked gives up while its own reply is still arriving.
The next query then flushes the port mid-line, and the fragment it
reads is a bare number: "LCE = 32" cut after the "=" is where a
diode current of 32 mA came from.
"""
deadline = time.monotonic() + self.timeout
while True:
cut = min((i for i in (self._rx.find(b'\r'), self._rx.find(b'\n'))
if i >= 0), default=-1)
if cut >= 0:
line = bytes(self._rx[:cut])
# CRLF is one terminator, not an empty line between two.
end = cut + (2 if self._rx[cut:cut + 2] == b'\r\n' else 1)
del self._rx[:end]
return line.decode('ascii', errors='replace').strip()
if time.monotonic() >= deadline:
return None
chunk = self.serial.read(self.serial.in_waiting or 1)
if not chunk:
return None # port timeout: nothing more is coming
self._rx += chunk
def _read_pending_lines(self) -> List[str]: def _read_pending_lines(self) -> List[str]:
"""Every further line the controller sends before the line goes quiet.""" """Every further line the controller sends before the line goes quiet."""
lines: List[str] = [] lines: List[str] = []
deadline = time.monotonic() + self.TRAILING_QUIET_S deadline = time.monotonic() + self.TRAILING_QUIET_S
while True: while True:
if self.serial.in_waiting: # What has already arrived is read whatever the quiet window
# says: the window is for deciding when to stop waiting, not
# for leaving a line in the buffer to confuse the next query.
if self._rx or self.serial.in_waiting:
line = self._read_line() line = self._read_line()
if line is None:
return lines # a partial line, nothing behind it
if line: if line:
lines.append(line) lines.append(line)
deadline = time.monotonic() + self.TRAILING_QUIET_S deadline = time.monotonic() + self.TRAILING_QUIET_S
@@ -175,7 +215,7 @@ class HeliosLaser:
mnemonic = fields[0].upper() if fields else "" mnemonic = fields[0].upper() if fields else ""
try: try:
# Anything volunteered while the port was idle answers no command. # Anything volunteered while the port was idle answers no command.
self.serial.reset_input_buffer() self._discard_input()
if not self._send_command(command): if not self._send_command(command):
return None return None
@@ -395,7 +435,7 @@ class HeliosLaser:
logger.error("Not connected to laser") logger.error("Not connected to laser")
return None return None
try: try:
self.serial.reset_input_buffer() self._discard_input()
if not self._send_command(command): if not self._send_command(command):
return None return None
first = self._read_line() first = self._read_line()
+87 -16
View File
@@ -12,23 +12,34 @@ import pytest
from hardware.helios_laser import HeliosLaser, PulseMode from hardware.helios_laser import HeliosLaser, PulseMode
# What the controller actually sends back, per Section 6 of the operator's # What the controller actually sends back, transcribed from a session with
# manual and the LER/LCE/CCE tables in Section 8. # the laser (tools/helios_lds_probe.py): CRLF line ends, the value padded
# out to a fixed width, and one or two blank lines closing every reply.
#
# b'LDS = 100 mA\r\n\r\n'
# b'LCE = 32\r\nBit 15..0: 0000 0000 0010 0000\r\n\r\n\r\n'
#
# The blank lines matter: a reader that stops at CR leaves the LF of the
# last one behind, and the next read waits out the port timeout for a CR
# that only the next command will bring.
_PAD = [""]
_REGISTER_PAD = ["", ""]
REPLIES = { REPLIES = {
"LDO": ["LDO = 1"], "LDO": ["LDO = 1 "] + _PAD,
"LDF": ["LDF = 20000 ns"], "LDF": ["LDF = 20000 ns"] + _PAD,
"LDS": ["LDS = 1500 mA"], "LDS": ["LDS = 1500 mA"] + _PAD,
"LDG": ["LDG = 14"], "LDG": ["LDG = 14 "] + _PAD,
"LRE": ["LRE = 0"], "LRE": ["LRE = 0 "] + _PAD,
"LTA": ["LTA = 25400 m°C"], "LTA": ["LTA = 25400 m°C"] + _PAD,
"LTT": ["LTT = 31200 m°C"], "LTT": ["LTT = 31200 m°C"] + _PAD,
"EOA": ["EOA = 40100 m°C"], "EOA": ["EOA = 40100 m°C"] + _PAD,
"CSR": ["CSR = 1234567"], "CSR": ["CSR = 1234567"] + _PAD,
"HSR": ["HSR = 7654321"], "HSR": ["HSR = 7654321"] + _PAD,
# The registers are the multi-line ones. # The registers are the multi-line ones.
"LER": ["LER = 0", "Bit 15..0: 0000 0000 0000 0000"], "LER": ["LER = 0", "Bit 15..0: 0000 0000 0000 0000"] + _REGISTER_PAD,
"LCE": ["LCE = 2", "Bit 15..0: 0000 0000 0000 0010"], "LCE": ["LCE = 2", "Bit 15..0: 0000 0000 0000 0010"] + _REGISTER_PAD,
"CCE": ["CCE = 0", "Bit 15..0: 0000 0000 0000 0000"], "CCE": ["CCE = 0", "Bit 15..0: 0000 0000 0000 0000"] + _REGISTER_PAD,
} }
@@ -182,7 +193,7 @@ def test_set_commands_clear_their_acknowledgement(laser):
def test_raw_command_returns_every_line(laser): def test_raw_command_returns_every_line(laser):
"""The diagnostics console is where a multi-line reply should be visible.""" """The diagnostics console is where a multi-line reply should be visible."""
assert laser.send_raw_command("LCE") == ( assert laser.send_raw_command("LCE") == (
"LCE = 2\nBit 15..0: 0000 0000 0000 0010" "LCE = 2\nBit 15..0: 0000 0000 0000 0010"
) )
@@ -254,3 +265,63 @@ def test_a_serial_number_still_comes_back_bare(laser):
laser.serial.replies = {"CSR": ["A1B2C3D4"], "HSR": ["7654321"]} laser.serial.replies = {"CSR": ["A1B2C3D4"], "HSR": ["7654321"]}
assert laser.get_controller_serial() == "A1B2C3D4" assert laser.get_controller_serial() == "A1B2C3D4"
assert laser.get_head_serial() == "7654321" assert laser.get_head_serial() == "7654321"
class SplitReplyPort(FakePort):
"""Answers LCE in two pieces, the tail arriving after the next command.
That is what the wire looks like when a query gives up early: at 9600
baud the rest of the reply is still coming, and reset_input_buffer()
cannot drop bytes that have not arrived. The fragment left over is
" 32" — the value half of "LCE = 32", which is a plausible
diode current and was read as one.
"""
def __init__(self):
super().__init__()
self._late = b""
def write(self, data: bytes) -> int:
text = data.decode("ascii").strip()
self.written.append(text)
mnemonic = text.split()[0].upper() if text.split() else ""
# Whatever is asked next, the last reply's tail lands in front of it.
self._buf += self._late
self._late = b""
if mnemonic == "LCE":
self._buf += b"LCE =" # ...and no line ending yet
self._late = b" 32\r\n\r\n\r\n"
return len(data)
for line in self.replies.get(mnemonic, []):
self._buf += line.encode("utf-8") + b"\r\n"
return len(data)
def test_a_late_fragment_is_not_the_next_query_s_value():
"""The regression this branch exists for.
LCE's reply is cut in half, so the register read gives up. The tail
arrives while the *next* query is being answered, and "32" is what the
panel showed as the pump diode current — LCE bit 5, "Door switch open",
read as milliamps.
"""
drv = HeliosLaser(timeout=1.0)
drv.serial = SplitReplyPort()
drv.is_connected = True
drv.TRAILING_QUIET_S = 0.0
assert drv._query_int("LCE") is None # cut off mid-reply
assert drv.get_current_ma() == 1500 # not 32
def test_a_reply_is_read_without_waiting_out_the_port(laser):
"""Nothing is left in either buffer once a reply has been read.
A leftover LF costs a whole port timeout on the next read, which is
what made a status poll take ~8.6 s against a 1 s interval.
"""
laser.timeout = 0.01 # a wait would show up as a failure below
assert laser.get_status_registers() == (0, 2, 0)
assert laser.get_current_ma() == 1500
assert laser.serial.in_waiting == 0
assert laser._rx == bytearray()
+3 -2
View File
@@ -43,8 +43,9 @@ REGISTERS = {"LER": LER_FLAGS, "LCE": LCE_FLAGS, "CCE": CCE_FLAGS}
# Reads that should not change anything, in the order the panel's poll # Reads that should not change anything, in the order the panel's poll
# issues them, plus the two the panel never asks for: LDG (pulse mode, which # issues them, plus the two the panel never asks for: LDG (pulse mode, which
# is what decides whether LDS is applied at all) and LMA (an actual-current # is what decides whether LDS is applied at all) and LMA — whose unit column
# read, per the manual's unit column). # in the manual says mA, but which this controller answers in m°C, so it is
# a resonator temperature and not a second current reading.
READ_SWEEP = ["LER", "LCE", "CCE", "LDO", "LDG", "LDF", "LDS", "LMA", "HTR"] READ_SWEEP = ["LER", "LCE", "CCE", "LDO", "LDG", "LDF", "LDS", "LMA", "HTR"]
QUIET_S = 0.4 # a reply is over once the line is idle this long QUIET_S = 0.4 # a reply is over once the line is idle this long