Auto-align: level the sample on the DC bias levels from the camera window
The operator frames a good spot, confirms the two DC levels the detector reads there, and the rig then measures its own tilt: step 1.5 mm either side on X and then on Y, and tilt the platform until those levels come back. The correction that fixes an offset point is the correction that levels the whole travel — height error and tilt effect are both proportional to the offset — so the procedure ends by applying it and leaving it applied. Both directions are measured from the same starting tilt and averaged, which makes their disagreement a flatness read-out rather than something averaged away silently. core/auto_align.py holds the geometry and the search, Qt-free. The three T-axes' azimuths are the whole geometry: T1 lies along +X so it alone tilts along X, and T0/T2 move as an equal-and-opposite pair to tilt along Y without touching X (tilt_response derives that, and the tests pin it — an axis map that drifts would still converge, on the wrong axis). The search is a secant null on the split-detector difference: probe once to learn what a microstep is worth, sign included, then step at the null. It refuses to servo on a scope that has not re-triggered, escalates a probe that reads as no response before calling an axis dead, and stops at a per-axis travel limit. gui/align_bridge.py runs it on a worker thread; stopping is a threading.Event rather than a queued command, because the worker is inside a long handler for the whole run. The camera window carries the button and the progress window, and locks the scan panel and the jog pads while a run owns the stage. Adds immediate MEAN measurements and an acquisition count to the scope driver, and read_bias_mv to core/scope_inspect — the one scalar the inspection state was missing. KNOWN_ISSUES.md records what only the rig can settle: the probe step, the travel limit, the hold current, and whether the piston the X phase applies alongside its tilt matters. Co-Authored-By: Claude Opus 5 <noreply@anthropic.com>
This commit is contained in:
@@ -64,3 +64,53 @@ in `configure_scan_trigger` is not enough for FastFrame to re-arm after an
|
||||
AVERAGE-mode sequence, and the row-packing warning ("N frames acquired, M
|
||||
expected") will say so in the log. Raise the settle rather than the ramp
|
||||
buffer — the stage geometry is not what changed.
|
||||
|
||||
## Auto-align: constants that are guesses until the rig confirms them
|
||||
|
||||
`core/auto_align.py` closes a loop over hardware whose gain nobody has
|
||||
measured. Three numbers in `AlignSettings`/`TAxisSettings` are reasoned
|
||||
defaults, not readings:
|
||||
|
||||
- `probe_steps = 200` — the first move of every search, made only to learn how
|
||||
many millivolts a microstep is worth. Too small and each search wastes
|
||||
iterations doubling it (the status line says so: "moved N microsteps and the
|
||||
DC difference did not change"); too large and the first move overshoots by
|
||||
more than the platform should be asked to travel in one go.
|
||||
- `max_excursion_steps = 20000` — the per-axis safety limit, measured from
|
||||
wherever the axis started. It exists to stop a runaway before the actuator
|
||||
reaches its end stop, so it has to be smaller than the real travel.
|
||||
- `hold_current_ma = 300` — the run current (600 mA) and microstepping (32)
|
||||
are specified; the standstill current is half the run current by analogy
|
||||
with the GR axis, and has not been checked against the platform's weight.
|
||||
|
||||
**Bench check:** run one auto-align and read the log. The first search's
|
||||
iteration count is the probe verdict — 3 or 4 steps means the probe is about
|
||||
right, and a "did not change by 2 mV" line means it is too small. Convert the
|
||||
applied corrections into actuator travel and compare against the T-axis
|
||||
travel to set the excursion limit. Watch whether the platform holds its tilt
|
||||
between the two phases; if it sags, raise the hold current.
|
||||
|
||||
## Auto-align: does the X phase's piston matter, and where is the pivot?
|
||||
|
||||
The X phase moves T1 alone, as specified. T1 is the only axis lying along X,
|
||||
so it does tilt the platform along X — but moving one leg of three also lifts
|
||||
the platform by a third of the move (`tilt_response(X_TILT)` returns a piston
|
||||
of 1/3 alongside the 2/3 tilt). The search nulls the split-detector
|
||||
difference, which a piston should not move, so the assumption is that the
|
||||
piston is harmless. The piston-free alternative is T1 +1 with T0 and T2 at
|
||||
−0.5 each.
|
||||
|
||||
Separately, the procedure assumes the tilt pivot is under the beam: if it is
|
||||
not, applying the correction shifts the DC levels at the reference point
|
||||
itself, and the Y phase then chases levels that no longer describe the rig.
|
||||
The code reports this rather than compensating for it — `AxisResult`'s
|
||||
"back at the reference" reading after the X phase is exactly that
|
||||
measurement.
|
||||
|
||||
**Bench check:** during an X search, watch DC1 + DC2 (the sum, not the
|
||||
difference) on the scope. If the sum moves as T1 moves, the piston is
|
||||
changing the amount of collected light and `X_TILT` should become the
|
||||
piston-free triple. Then read the X phase's reference residual out of the
|
||||
log: more than a few millivolts means the pivot is not under the beam, and
|
||||
the Y phase's reference should be re-measured after the X correction instead
|
||||
of reusing the operator's original numbers.
|
||||
|
||||